A determination of lead content in plant by microwaveheating and GFAAS
-
-
Abstract
The best experiment conditions, accuracy and precision for determining the microamounts of lead content in plant by microwaveheating GFAAS approach were studied. Results indicated that the best acid system is 5 mL HNO33 mL H2O22 mL HF, the best microwaveheating condition is under the state of power 600 W、temperature 180 ℃ and microwaveheating 15 min.The optimal working condition of Graphite Furnace is the dried 45 s under 90 ℃, cineration 20 s under 900 ℃, atomized temperature 1 400 ℃ and cleaned temperature 2 500 ℃.The limit of detection of this method was found to be 0.068 4 mg/kg,the relative standard deviation error were in the range of 1.2%-5.6%,and recoveries were in the range of 95.81%-100.09%.
-
-